Special Features:
- Optimum operational reliability
- Universal applicability
- Adaption to nearly all process condi- tions due to its compact and modular design
- Easy mounting and maintenance
- Small volume, fast response time
Technical Data:
Gas Sample Probe Version | SP2000 | SP2000-H/R | SP2300-H | SP2400-H |
Part No. | 20S1000 | 20S2000 | 20S3000 | 20S3500 |
Protective cover | no | yes | yes | yes |
Degree of protection | IP54 EN60529 | |||
Filter housing material | Stainless steeel 316* | PTFE | Titanium | |
Sealing materials | FPM* /7aT** = PTFE /H320** = graphite | |||
Probe flange sealing material | Novapress | |||
Insitu probe tube/prefilter | Optional | |||
Sample pressure max. | 0,4-6 bar* abs., /7aT**= 2 bar abs., /HP** = 25 bar abs. | 0,4-2 bar abs. | 0,4-6 bar abs. | |
Ambient temperature | 20 °C to +180 °C-20 °C to +60 °C*** /PT100, /Fe-CuNi, /Ni-CrNi** = -20°C to +80°C | |||
Filter chamber volume | 120 cm3 | |||
Filter element, porosity | S-2K150= Ceramic*, 2micron, /F-0, 1GF150= Glass fibre**, 0,1 micron, /FW= Spun glass** | |||
Thermostat, Temperature adjustment | 0-180°C* /H320**= 0-320°C /PT100** /Fe-CuNi** /Ni-CrNi** | |||
Ready for operation | after 40min /H320** = after 60min | |||
Low temperature alarm contact* | G1/4″ i DIN ISO 228/1 | |||
Sample gas outlet connection | 1x 1/4“ NPTi* tube connectors ø 6, 8 or 10 mm** /H320**= 6 mm | |||
Backflush/Test gas connection | 1/4“ NPTi* /R**, /H320**= tube ø 6 mm | |||
Power supply | 230V 50/60 Hz, 800W /115V** = 115V 60Hz, 800W (fuse protection 10A) | |||
Electrical connections | Terminals max. 4mm2, 2x PG 13,5 cable gland | |||
Electrical equipment standard | EN 61010, EN 60519-1 | |||
Mounting flange | DN65 PN6-B >DN or ANSI possible** /HP** = DN50 PN25 | |||
Mounting flange material | SS316Ti | PTFE | Titanium | |
Weight | 7 kg* | 15,4 kg* | 15,4 kg* | 14,5kg* |
*Standard
**Options (/H320 not for SP2300-H, /7aT** not for SP2300-H and SP2400-H)
***In case of higher ambient temperatures use option PT100 (Part No. 20S9025) or thermocouple Fe-CuNi
respectively Ni-CrNi (Part No. 20S9027 resp. 20S9028) instead of the thermo-state controller.
Then, an additional electronic temperature controller is necessary.