Special Features:

  • Completely pre-installed and opera- tional within 10 minutes
  • Compact construction, takes up little space
  • Self-monitoring and low maintenance
  • Up to 5 calibration gases manually or externally switchable
  • Outlet dew point adjustable from +2 °C to +15 °C
  • Dew point stability ± 0,1 °C
  • Optimum reliability

Technical Data:

19″-Gas Conditioning System Series CSS® version CSS version CSS-3
Inlets and max. gas flow** 1x 140 Nl/hr 1x 350 Nl/hr
Outlets incl. flowmeters 2x à 70 Nl/hr 2x à 250 Nl/hr
Gas pressure 0,7 to 1,2 bar abs.
Sample gas inlet temperature** max. 150 °C
Sample gas inlet dewpoint** max. 80 °C
Sample gas outlet dewpoint range of adjustment: +2 °C. +15 °C, factory setting: +5 °C
Dew point stability at constant conditions < ±0,1 °C
Ambient temperature** +5 °C to +40 °C +5 °C to +45 °C
Total cooling capacity at +25 °C ambient temperature max. 50 kJ/hr max. 90 kJ/hr
Storage temperature -25 °C to +65 °C
Relative humiditiy <80%
Ready for work approx. 10 min
Gas filter glassfibre, porosity 0,1 µm
Housing 19″ rack mounting 6U, depth 350 mm, color RAL 7032
Degree of protection IP 20-EN 60529
Tube connections G 1/4 i -DIN ISO 228/1*
Power supply 230V, 48-62Hz, 150VA or

115V, 48-62Hz, 150VA

230V, 48-62Hz, 250VA or

115V, 48-62Hz, 250VA

Electrical equipment standard EN 61010
Electrical connections power terminals max 4 mm2, 4x PG13,5

Alarm- / control signals 15-pin Sub-D connector

Status signal cooler temperature, liquid alarm, flow alarm
Satus alarm contact, contact rating potential free changeover contact, max. 24V, 1A
Test gas inlets 2 solenoid valves, actuated manually orby external switch* (optional +3 pcs)
Material of sample contacting parts PVDF, glass, FPM, Novoprene®, PTFE
Weight approx. 15 kg approx. 16,5 kg
Option:
Electronic PID temperature contoller 703 for heated sample lines front surface mounting, range 0-200 °C, sensor PT 100 or Fe-CuNi,

controlling outlet with a solid state relais 10A # alarm output integrated into the CSS status alarm,

parameter free configurable

 

Equipment conditioning unit CSS conditioning unit CSS-3
Sample gas inlet 1
Sample gas outlets 2
Flowmeter with needle valve FM 40 70 Nl/hr air FM 40 250 Nl/hr air
Electro gas-cooler ECP 1000-G90° for 150 Nl/hr ECP 3000-G90° for 350 Nl/hr
Diaphragm pump N3KPE N9KPE
Peristaltic pump SR 25.1
Micro fine filter FPF 0,1 GF with glass-fibre filter element 0,1 µm
External or manual switching to 1x 3/2-way solenoid valve 124C3 and
testing with zero or span gas 2x 2/2-way solenoid valves 2011, optional +3 pcs.
Hosing and connectors PTFE-PVDF
Status alarm with optical display for cooler temperature, flow failure and liquid alarm
Optional: electronic temperature controller for heated sample line