Special Features:

  • Optimum operational reliability
  • Universal applicability
  • Adaption to nearly all process condi- tions due to its compact and modular design
  • Easy mounting and maintenance
  • Small volume, fast response time

Technical Data:

Gas Sample Probe Version SP2000 SP2000-H/R SP2300-H SP2400-H
Part No. 20S1000 20S2000 20S3000 20S3500
Protective cover no yes yes yes
Degree of protection IP54 EN60529
Filter housing material Stainless steeel 316* PTFE Titanium
Sealing materials FPM* /7aT** = PTFE /H320** = graphite
Probe flange sealing material Novapress
Insitu probe tube/prefilter Optional
Sample pressure max. 0,4-6 bar* abs., /7aT**= 2 bar abs., /HP** = 25 bar abs. 0,4-2 bar abs. 0,4-6 bar abs.
Ambient temperature 20 °C to +180 °C-20 °C to +60 °C*** /PT100, /Fe-CuNi, /Ni-CrNi** = -20°C to +80°C
Filter chamber volume 120 cm3
Filter element, porosity S-2K150= Ceramic*, 2micron, /F-0, 1GF150= Glass fibre**, 0,1 micron, /FW= Spun glass**
Thermostat, Temperature adjustment 0-180°C* /H320**= 0-320°C /PT100** /Fe-CuNi** /Ni-CrNi**
Ready for operation after 40min /H320** = after 60min
Low temperature alarm contact* G1/4″ i DIN ISO 228/1
Sample gas outlet connection 1x 1/4“ NPTi* tube connectors ø 6, 8 or 10 mm** /H320**= 6 mm
Backflush/Test gas connection 1/4“ NPTi* /R**, /H320**= tube ø 6 mm
Power supply 230V 50/60 Hz, 800W /115V** = 115V 60Hz, 800W (fuse protection 10A)
Electrical connections Terminals max. 4mm2, 2x PG 13,5 cable gland
Electrical equipment standard EN 61010, EN 60519-1
Mounting flange DN65 PN6-B >DN or ANSI possible** /HP** = DN50 PN25
Mounting flange material SS316Ti PTFE Titanium
Weight 7 kg* 15,4 kg* 15,4 kg* 14,5kg*

*Standard
**Options (/H320 not for SP2300-H, /7aT** not for SP2300-H and SP2400-H)
***In case of higher ambient temperatures use option PT100 (Part No. 20S9025) or thermocouple Fe-CuNi
respectively Ni-CrNi (Part No. 20S9027 resp. 20S9028) instead of the thermo-state controller.
Then, an additional electronic temperature controller is necessary.