Special Features:

  • Ambient temperature from +5 to +50 °C[41 to 122 °F] (no extra charge)
  • Outlet dew point adjustable from+2 to +15 °C [35.6 °F to 59 °F]
  • Dew point stability ± 0.1 °C [± 0.18 °F]
  • Control set point selectable between:
    absolute value mode or ∆T mode
  • Optional measurement of the gas outlet
    temperature in the gas path
  • Ready for use in less than 3 minutes
  • Option: configurable mA output (no
    shielding necessary)

 

  • Integrated evaluation for liquid alarm
    sensors
  • Configurable, potential-free alarm
    output
  • Universal power supply
  • Mounting option for up to two SR25.2W
    peristaltic pumps below the unit
  • Compact design, low weight
  • Jet-Stream heat exchangers available in
    various materials
  • Compatible with previous version

Technical Data:

Gas Cooler Serie EC® ECP1000C ECP2000C ECP3000C
Part number without heat exchangers 01 K 1400x 01 K 2400x 01 K 3400x
Number of possible heat exchangers 1 2 1
Gas flow rate per heat exchanger max. 150 Nl/h** 2 x max. 150 Nl/h** max. 350 Nl/h**
Ambient temperature +5 to +50 °C [41 to 122 °F]
Storage temperature -20 to +60 °C [~-4 to 140 °F]
Sample outlet dew point range of adjustment: +2 to +15 °C [35.6 to 59 °F], factory setting: +5 °C [41 °F]
Dew point stability ±0.1 °C [±0.18 °F] at constant conditions
Sample inlet temperature max. 180 °C [356 °F]**
Sample inlet dew point max. 80 °C [176 °F]**
Total cooling power at +25 °C ambient 110 kJ/h 2 x 90 kJ/h 110 kJ/h
∆ P per heat exchanger at 1 mbar at 150 Nl/h 1 mbar at 150 Nl/h 5 mbar at 350 Nl/h
Stagnant space per heat exchanger 50 ml 2 x 50 ml 100 ml
Power consumption 150 VA 275 VA 150 VA
Power supply 115 – 230 V AC ±10 %, 50/60 Hz
Ready for use <3 min. (at 25 °C ambient temperature and with no load applied)
Max. loudness 58 dBA
Electrical connection Power: Pluggable via solenoid valve plug type A
Alarm relay: Pluggable via solenoid valve plug type B
mA: When purchasing the mA option, pluggable via Phoenix circular connector 1681101
Signal input and output One mA output per channel possible (no shielding required)
One M&C LA liquid alarm sensor can be connected per channel. The evaluation is integrated as standard.
Status alarm: 2 changeover contacts Contact rating: 250 V, 2 A, 500 VA, 50 W
Case protection IP 20; EN 60529
Electrical standard EN 61010
EMV standard EN 61326
Case color RAL 9003
Method of mounting wall-mount
Case dimensions (W x H x D) 300 x 200 x 225 mm [11.8” x 7.9” x 8.9”]
Weight without heat exchangers 6.5 kg [14.3 lbs.] 8.2 kg [18.1 lbs.] 6.7 kg [14.8 lbs.]

* Optional
** Maximum values in technical data must be rated in consideration of total cooling capacity at 25 °C [77 °F]
ambient temperature and an outlet dew point of 5 °C [41 °F].

Gas Cooler Serie ECP® ECP1000C/ ECP2000C
Heat exchanger type ECM-2/ ECP(1/2)000C/ ECC-1 G, WT ECM-2/ ECP(1/2)000C/ ECC-1 PV, WT ECM-2/ ECP(1/2)000C/ ECC-1 SS, WT ECM-2/ ECP(1/2)000C/ ECC-1 SS/NPT, WT ECM-2/ ECP(1/2)000C/ ECC-1 G/GL14, WT
Part.-No. 97K0100 97K0110 97K0115 97K0115NN 97K0101
Material of heat exchanger Duranglas® PVDF SS 316Ti SS 316Ti Duranglas®
Sample gas connection GL18 for tube Ø 6 mm OD Tube Ø 6 mm Tube Ø 6 mm 1/4” tube GL18 for tube Ø 6 mm OD; GL14 for sensor
Condensate connec- tion GL25 for tube Ø 12 mm, Ø 8 mm* or Ø 10 mm* G 3/8’’ i G 3/8’’ i 3/8“NPT GL25 for tube Ø 12 mm, Ø 8 mm* or Ø 10 mm*

 

Gas Cooler Serie ECP® ECP3000C
Heat exchanger type ECM-1/ECP3000(C)/ ECC-1 G, WT ECM-1/ECP3000(C)/ ECC-1 PV, WT ECM-1/ECP3000(C)/ ECC-1 SS, WT ECM-1/ECP3000(C)/ ECC-1 SS/NPT, WT
Part.-No. 97K0140 97K0170 97K0160 97K0160N
Material of heat exchanger Duranglas® PVDF SS 316Ti SS 316Ti
Sample gas connection GL18 for tube Ø 6 mm OD G 1/4’’ i G 1/4’’ i 1/4“ NPT
Condensate connection GL25 for tube Ø 12 mm, Ø 8 mm* or Ø 10 mm* G 3/8’’ i G 3/8’’ i 3/8“ NPT

* Option:-
1)With GL adapter.
2)With peristaltic pump SR25.2W